Information
Infrared pyrometers are frequently used to monitor and control manufacturing processes. The HS-25-400 calibration radiator is used as part of measurement equipment monitoring to carry out regular checks on pyrometers. Thanks to its compact design, it is suitable for installation in machinery and plant. The blackbody can be used to check individual operating points or to carry out complete recalibrations.
The calibration of the pyrometers is supported by the LASCON Process Manager software and can also be carried out easily by the user.
- Calibration radiator:
- Temperature range: 100°C to 400°C (non-cooling)
- Emissivity: 0.97 in the range 1.5 µm – 2.2 µm
- Heating element dimensions: diameter = 25 mm
- Heat output: 100 W, 24 V AC
- Heating rate: 50°C/min
- Settling time 20°C to 200°C: 4 min
- Accuracy: ±2°C plus ±1% of the reading
- Long-term accuracy: ±0.1°C at 200°C after a 4-minute dwell time
- Display resolution: 0.1°C
- Dimensions (W x H x D): 80 x 43 x 110 mm (cable connector not plugged in)
- Weight: 0.7 kg
- Control unit:
- Supply voltage: 230 V AC, 50 Hz (115 V AC on request)
- Current consumption: max. 0.3 A
- Overcurrent protection: 2 x 5 x 20 mm 0.8 A micro-fuses
- Safety features: sensor break detection, sensor short-circuit detection
- Overtemperature limit
- Option: RS485 communication
- Ambient temperature: 5°C to 40°C
- Weight: approx. 2 kg (including mains cable)
- Dimensions (WxHxD): 60 x 105 x 240 mm (cable not connected, controller module
- protruding 10 mm above the front panel)